Locked minimally invasive plating versus fourth generation nailing in the treatment of AO/OTA 31A2.2 fractures: A biomechanical comparison of PCCP(®) and Intertan nail(®).
- Source :
-
Injury [Injury] 2015 Aug; Vol. 46 (8), pp. 1475-82. Date of Electronic Publication: 2015 May 14. - Publication Year :
- 2015
-
Abstract
- Introduction: Locked minimally invasive plating and fourth generation nailing potentially could reduce the complication rate in the treatment of trochanteric femur fractures by its rotational stability and providing better lateral cortical support. The purpose of this study was (1) to compare the biomechanical properties of the Percutaneous compression plate (PCCP) and the Intertan nail (IT) with regards to implant failure and (2) to assess dynamic stability coefficients in an unstable AO/OTA 31A2.2 fracture model.<br />Methods: In paired femurs, a standardised unstable trochanteric femur fracture was induced by an oscillating saw. The fractures were stabilised by either the PCCP (Orthofix, McKinney, TX, USA) or the IT (Smith & Nephew, Memphis, TN, USA). All femurs were loaded with 300N, followed by an increase in load until failure using 300N each time (2000 cycles each, 0.5Hz). After every load step the samples were assessed visually and radiographically. We measured migration and performed a survival analysis.<br />Results: 16 fractures were induced in 8 paired human specimens (mean age: 84 years, 61-100 years). The mean stiffness (PCCP vs. IT: 249±124N/mm vs. 273±153N/mm; p=0.737) was comparable. The IT proved superior to the PCCP with regard to the number of cycles reached before failure occurred (PCCP vs. IT: 12,691±4733 vs. 15,313±4875 cycles; p=0.023). Except for a higher axial migration of the IT at failure point (PCCP vs. IT: 1.3mm vs. 4.3mm; p=0.028) there were no differences between the intra- and extramedullary implants, not even in terms of rotational stability along the femoral neck axis. A fracture of the femoral neck caused test abortion in both implants in most cases.<br />Conclusion: This study showed a superiority of the IT compared with the PCCP with regards to number of cycles achieved under sequential load increases for unstable trochanteric femur fractures. The stiffness was comparable. Both implants showed a high rotational stability and a support of the lateral wall.<br />Study Type: Biomechanical study.<br /> (Copyright © 2015 Elsevier Ltd. All rights reserved.)
- Subjects :
- Aged
Aged, 80 and over
Biomechanical Phenomena
Cadaver
Equipment Failure Analysis
Female
Humans
Male
Prosthesis Design
Tensile Strength
Bone Nails
Bone Plates
Femoral Fractures surgery
Fracture Fixation, Internal instrumentation
Fracture Fixation, Intramedullary instrumentation
Minimally Invasive Surgical Procedures instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 1879-0267
- Volume :
- 46
- Issue :
- 8
- Database :
- MEDLINE
- Journal :
- Injury
- Publication Type :
- Academic Journal
- Accession number :
- 25997559
- Full Text :
- https://doi.org/10.1016/j.injury.2015.05.011