Back to Search Start Over

Sources of variability in the quantification of tissue optical properties by multidiameter single-fiber reflectance and fluorescence spectroscopy.

Authors :
Brooks S
Hoy CL
Amelink A
Robinson DJ
Nijsten TE
Source :
Journal of biomedical optics [J Biomed Opt] 2015 May; Vol. 20 (5), pp. 57002.
Publication Year :
2015

Abstract

Recently, a multidiameter single-fiber reflectance and fluorescence spectroscopy device has been developed that enabled us to extract the autofluorescence of tissue that is corrected for the optical properties. Such a system has been incorporated in the population-based Rotterdam Study to investigate the autofluorescence of the skin. Since the device will be used by different operators over many years, it is essential that the results are comparable between users. It is, however, unclear how different methods of handling the probe might influence the outcome. Variability of blood oxygen saturation, blood volume fraction and vessel diameter, average gamma, reduced scattering coefficient at 800 nm, and integrated intrinsic fluorescence measured in three volunteers were assessed within and between eight untrained users. A variability of less than one standard deviation from the group mean was defined as an acceptable limit. Three mature volunteers were also included to assess the intrauser variability of mature skin. The variation in the measured parameters suggests that variation is dominated by tissue heterogeneity. Most users measured within one standard deviation of the group mean. Notably, corrected intrinsic fluorescence showed low intra- and interuser variability. These results strongly suggest that variability is mostly caused by tissue heterogeneity and is not user induced.

Details

Language :
English
ISSN :
1560-2281
Volume :
20
Issue :
5
Database :
MEDLINE
Journal :
Journal of biomedical optics
Publication Type :
Academic Journal
Accession number :
25938208
Full Text :
https://doi.org/10.1117/1.JBO.20.5.057002