Inadvertent Detachment of a Retrievable Intracranial Stent: Review of Manufacturer and User Facility Device Experience.
- Source :
-
The neuroradiology journal [Neuroradiol J] 2015 Apr; Vol. 28 (2), pp. 172-6. Date of Electronic Publication: 2015 Apr 28. - Publication Year :
- 2015
-
Abstract
- Few systematic surveys have dealt with the potential procedural risks associated with the use of retrievable intracranial stents [Solitaire Flow Restoration (Solitaire FR)], which have become effective tools for recanalizing acutely occluded cerebral arteries. The aim of this study was to present the real-world experiences of Solitaire-FR-related adverse events by reviewing the MAUDE (Manufacturer and User Facility Device Experience) as published on the United States Food and Drug Administration website. In total, 85 adverse events related to the use of the Solitaire FR stent were reported between March 2012 and October 2014. In 80 patients these adverse events were attributable to inadvertent detachment of the device. Thirteen of these 80 patients (16%) died after the procedure. Morbidity data were available in 62 patients, among whom 11 (18%) had suffered a procedure-related injury. Detachment occurred at the first, second, and third pass in nine (21%), 21 (49%), and 13 (30%) of the 43 patients for whom this information was available, respectively. Resistance was perceived by the physician during retrieval of the device in 12 patients, and lesion characteristics were noted in 13. A rescue maneuver was reported in 20 (25%) of the 80 patients in whom the adverse event was attributable to detachment of the device, resulting in flow reestablishment in 13 (65%). The risk of inadvertent detachment during stent retrieval cannot be overemphasized in real-world scenarios, and careful consideration of the "dos and don'ts" is essential for the achievement of a safe procedure.<br /> (© The Author(s) 2015 Reprints and permissions: sagepub.co.uk/journalsPermissions.nav.)
- Subjects :
- Causality
Cerebral Revascularization methods
Device Removal instrumentation
Endovascular Procedures instrumentation
Equipment Failure Analysis
Foreign-Body Migration etiology
Humans
Incidence
Intracranial Thrombosis epidemiology
Intracranial Thrombosis etiology
Intracranial Thrombosis therapy
Mechanical Thrombolysis instrumentation
Republic of Korea epidemiology
Retrospective Studies
Risk Factors
Stents statistics & numerical data
Stroke prevention & control
Survival Rate
Vascular Calcification epidemiology
Vascular Calcification etiology
Vascular Calcification therapy
Vascular System Injuries etiology
Vascular System Injuries therapy
Cerebral Arteries injuries
Cerebral Revascularization instrumentation
Equipment Failure statistics & numerical data
Foreign-Body Migration epidemiology
Stents adverse effects
Vascular System Injuries epidemiology
Subjects
Details
- Language :
- English
- ISSN :
- 1971-4009
- Volume :
- 28
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- The neuroradiology journal
- Publication Type :
- Academic Journal
- Accession number :
- 25923678
- Full Text :
- https://doi.org/10.1177/1971400915576650