Back to Search
Start Over
Comparative evaluation of two matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems, Vitek MS and Microflex LT, for the identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories.
- Source :
-
Journal of microbiological methods [J Microbiol Methods] 2015 Jun; Vol. 113, pp. 13-5. Date of Electronic Publication: 2015 Mar 25. - Publication Year :
- 2015
-
Abstract
- We evaluated the performance of two MALDI-TOF MS systems for the identification of clinically important Gram-positive cocci. Vitek MS and Microflex LT correctly identified 97.2% and 94.7%, respectively. Both systems offer reliable and rapid identification of clinically important Gram-positive cocci isolated in clinical laboratories, including staphylococci, streptococci, and enterococci. Expanding the databases, especially of coagulase-negative staphylococci and viridans streptococci, would enhance performance.<br /> (Copyright © 2015 Elsevier B.V. All rights reserved.)