Back to Search Start Over

Continuous monitoring of regional function by a miniaturized ultrasound transducer allows early quantification of low-grade myocardial ischemia.

Authors :
Hyler S
Pischke SE
Halvorsen PS
Espinoza A
Bergsland J
Tønnessen TI
Fosse E
Skulstad H
Source :
Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography [J Am Soc Echocardiogr] 2015 Apr; Vol. 28 (4), pp. 486-94. Date of Electronic Publication: 2015 Jan 27.
Publication Year :
2015

Abstract

Background: Sensitive methods for the early detection of myocardial dysfunction are still needed, as ischemia is a leading cause of decreased ventricular function during and after heart surgery. The aim of this study was to test the hypothesis that low-grade ischemia could be detected quantitatively by a miniaturized epicardial ultrasound transducer (Ø = 3 mm), allowing continuous monitoring.<br />Methods: In 10 pigs, transducers were positioned in the left anterior descending and circumflex coronary artery areas. Left ventricular pressure was obtained by a micromanometer. The left internal mammary artery was grafted to the left anterior descending coronary artery, which was occluded proximal to the anastomosis. Left internal mammary artery flow was stepwise reduced by 25%, 50%, and 75% for 18 min each. From the transducers, M-mode traces were obtained, allowing continuous tissue velocity traces and displacement measurements. Regional work was assessed as left ventricular pressure-displacement loop area. Tissue lactate measured from intramyocardial microdialysis was used as reference method to detect ischemia.<br />Results: All steps of coronary flow reduction demonstrated reduced peak systolic velocity (P < .05) and regional work (P < .01).The decreases in peak systolic velocity and regional work were closely related to the degree of ischemia, demonstrated by their correlations with lactate (R = -0.74, P < .01, and R = -0.64, P < .01, respectively). The circumflex coronary artery area was not affected by any of the interventions.<br />Conclusions: The epicardially attached miniaturized ultrasound transducer allowed the precise detection of different levels of coronary flow reduction. The results also showed a quantitative and linear relationship among coronary flow, ischemia, and myocardial function. Thus, the ultrasound transducer has the potential to improve the monitoring of myocardial ischemia and to detect graft failure during and after heart surgery.<br /> (Copyright © 2015 American Society of Echocardiography. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1097-6795
Volume :
28
Issue :
4
Database :
MEDLINE
Journal :
Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography
Publication Type :
Academic Journal
Accession number :
25636368
Full Text :
https://doi.org/10.1016/j.echo.2014.12.014