Back to Search Start Over

Hemodynamic changes and retrograde flow in LVAD failure.

Authors :
Giridharan GA
Koenig SC
Soucy KG
Choi Y
Pirbodaghi T
Bartoli CR
Monreal G
Sobieski MA
Schumer E
Cheng A
Slaughter MS
Source :
ASAIO journal (American Society for Artificial Internal Organs : 1992) [ASAIO J] 2015 May-Jun; Vol. 61 (3), pp. 282-91.
Publication Year :
2015

Abstract

In the event of left ventricular assist device (LVAD) failure, we hypothesized that rotary blood pumps will experience significant retrograde flow and induce adverse physiologic responses. Catastrophic LVAD failure was investigated in computer simulation with pulsatile, axial, and centrifugal LVAD, mock flow loop with pulsatile (PVAD) and centrifugal (ROTAFLOW), and healthy and chronic ischemic heart failure bovine models with pulsatile (PVAD), axial (HeartMate II), and centrifugal (HVAD) pumps. Simulated conditions were LVAD "off" with outflow graft clamped (baseline), LVAD "off" with outflow graft unclamped (LVAD failure), and LVAD "on" (5 L/min). Hemodynamics (aortic and ventricular blood pressures, LVAD flow, and left ventricular volume), echocardiography (cardiac volumes), and end-organ perfusion (regional blood flow microspheres) were measured and analyzed. Retrograde flow was observed with axial and centrifugal rotary pumps during LVAD failure in computer simulation (axial = -3.4 L/min, centrifugal = -2.8 L/min), mock circulation (pulsatile = -0.1 L/min, centrifugal = -2.7 L/min), healthy (pulsatile = -1.2 ± 0.3 L/min, axial = -2.2 ± 0.2 L/min, centrifugal = -1.9 ± 0.3 L/min), and ischemic heart failure (centrifugal = 2.2 ± 0.7 L/min) bovine models for all test conditions (p < 0.05). Differences between axial and centrifugal LVAD were statistically indiscernible. Retrograde flow increased ventricular end-systolic and end-diastolic volumes and workload, and decreased myocardial and end-organ perfusion during LVAD failure compared with baseline, LVAD support, and pulsatile LVAD failure.

Details

Language :
English
ISSN :
1538-943X
Volume :
61
Issue :
3
Database :
MEDLINE
Journal :
ASAIO journal (American Society for Artificial Internal Organs : 1992)
Publication Type :
Academic Journal
Accession number :
25635935
Full Text :
https://doi.org/10.1097/MAT.0000000000000200