Back to Search Start Over

Reconstruction of 2D x-ray radiographs at the National Ignition Facility using pinhole tomography (invited).

Authors :
Field JE
Rygg JR
Barrios MA
Benedetti LR
Döppner T
Izumi N
Jones O
Khan SF
Ma T
Nagel SR
Pak A
Tommasini R
Bradley DK
Town RP
Source :
The Review of scientific instruments [Rev Sci Instrum] 2014 Nov; Vol. 85 (11), pp. 11E503.
Publication Year :
2014

Abstract

Two-dimensional radiographs of imploding fusion capsules are obtained at the National Ignition Facility by projection through a pinhole array onto a time-gated framing camera. Parallax among images in the image array makes it possible to distinguish contributions from the capsule and from the backlighter, permitting correction of backlighter non-uniformities within the capsule radiograph. Furthermore, precise determination of the imaging system geometry and implosion velocity enables combination of multiple images to reduce signal-to-noise and discover new capsule features.

Details

Language :
English
ISSN :
1089-7623
Volume :
85
Issue :
11
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
25430345
Full Text :
https://doi.org/10.1063/1.4890395