Back to Search Start Over

Capacitance-to-digital: the upgrades of single chip detector.

Authors :
Drevinskas T
Maruška A
Briedis V
Source :
Electrophoresis [Electrophoresis] 2015 Jan; Vol. 36 (2), pp. 292-7. Date of Electronic Publication: 2014 Nov 10.
Publication Year :
2015

Abstract

The capacitance-to-digital single chip detector was upgraded. The paper discusses hardware issues and benefits of the designed/upgraded detector. The device can be operated from rechargeable lithium-ion battery as stand-alone, portable system and is capable of transmitting real-time data wirelessly. The detector and additional modules (battery, battery holder, microcontroller board, wireless module) weight is less than 85 g. Electrophoretic separation in low conductivity 20 mM MES/L-His buffer, pH 6.1, was performed in order to evaluate detection parameters. The system is capable of quantification of potassium ions down to 0.31 μM. Investigation of differential signal acquisition configuration showed improved performance regarding external noise and temperature fluctuations. The system can be a solution for stand-alone, field-portable capillary format separation detector.<br /> (© 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.)

Details

Language :
English
ISSN :
1522-2683
Volume :
36
Issue :
2
Database :
MEDLINE
Journal :
Electrophoresis
Publication Type :
Academic Journal
Accession number :
25288122
Full Text :
https://doi.org/10.1002/elps.201400320