Back to Search
Start Over
Higher contact-force values associated with better mid-term outcome of paroxysmal atrial fibrillation ablation using the SmartTouch™ catheter.
- Source :
-
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2015 Jan; Vol. 17 (1), pp. 56-63. Date of Electronic Publication: 2014 Oct 03. - Publication Year :
- 2015
-
Abstract
- Aims: Real-time measurement of contact force (CF) during catheter ablation of atrial fibrillation (AF) has been recently suggested to potentially impact procedural outcome. However, the role of CF intensity on mid-term results using the SmartTouch™ catheter has not been investigated so far.<br />Methods and Results: Pulmonary vein isolation (PVI) using the SmartTouch™ catheter was performed in 100 eligible patients (age 62 ± 8; 79% men) undergoing a first procedure of paroxysmal AF catheter ablation. Continuous CF monitoring during catheter ablation allowed calculation of mean CF per patient. Patients were dichotomized into high CF (≥22 g, upper quartile) and low CF (<22 g, remaining) and enroled in a standardized follow-up programme (after a 3-month blanking period), free from antiarrhythmic therapy, with regular evaluations including 24 h Holter recordings at 1, 3, 6, 9, 12, 18, and 24 months. Atrial fibrillation relapse was defined as any symptomatic or asymptomatic atrial arrhythmia lasting >30 s. The average CF among all procedures was 19.6 ± 3.7 g. Though complete PVI was eventually achieved in all cases in both groups, success using an exclusively anatomical approach was higher in the high CF group (92.0 vs. 72.0%; P = 0.04). During a mean follow-up of 19 ± 5 months, a lower incidence of AF relapse was observed in higher CF patients (4.0 vs. 20.0%; log rank P = 0.04). Pericardial tamponade occurred in one patient in the higher CF group. No thromboembolism or procedure-associated deaths were observed.<br />Conclusion: Higher values of CF overall during antral PVI appear to be associated with a higher likelihood of sinus rhythm maintenance without significantly increasing the complication rate.<br /> (Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email: journals.permissions@oup.com.)
- Subjects :
- Catheter Ablation adverse effects
Catheter Ablation methods
Equipment Design
Equipment Failure Analysis
Female
Humans
Male
Middle Aged
Stress, Mechanical
Surgery, Computer-Assisted adverse effects
Touch
Treatment Outcome
User-Computer Interface
Atrial Fibrillation diagnosis
Atrial Fibrillation surgery
Body Surface Potential Mapping instrumentation
Catheter Ablation instrumentation
Surgery, Computer-Assisted instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 1532-2092
- Volume :
- 17
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
- Publication Type :
- Academic Journal
- Accession number :
- 25280910
- Full Text :
- https://doi.org/10.1093/europace/euu218