Back to Search Start Over

Mechanical properties and in vivo performance of load-bearing fiber-reinforced composite intramedullary nails with improved torsional strength.

Authors :
Moritz N
Strandberg N
Zhao DS
Mattila R
Paracchini L
Vallittu PK
Aro HT
Source :
Journal of the mechanical behavior of biomedical materials [J Mech Behav Biomed Mater] 2014 Dec; Vol. 40, pp. 127-139. Date of Electronic Publication: 2014 Sep 01.
Publication Year :
2014

Abstract

Fiber-reinforced composites (FRC) could be feasible materials for fracture fixation devices if the mechanical properties of the composites are congruent with the local structural properties of bone. In a recently developed FRC implant, bisphenol A dimethacrylate (BisGMA) and triethylene glycol dimethacrylate (TEGDMA) resin was reinforced with unidirectional E-glass fibers. The addition of a braided glass fiber sleeving to the unidirectional fibers increased the torsional strength (99.5MPa) of the FRC implants at the expense of the flexural strength (602.0MPa). The flexural modulus was 15.3GPa. Two types of FRC intramedullary nails were prepared; first type was FRC as such, second type was FRC with a surface layer of bioactive glass (BG) granules. Experimental oblong subtrochanteric defect was created in 14 rabbits. The defect, which reduced the torsional strength of the bones by 66%, was fixed with an FRC intramedullary nail of either type. The contralateral intact femur served as the control. This model simulated surgical stabilization of bone metastasis. After 12 weeks of follow-up, the femurs were harvested and analyzed by torsional testing, micro-CT and hard tissue histology. Healed undisplaced peri-implant fractures were noticed in half of the animals irrespective of the type of FRC implant. Torsional testing showed no significant differences between the implantation groups. The torsional strength of the bones stabilized by either type of FRC implant was 83% of that of the contralateral femurs. In histological analysis, no implant debris and no adverse tissue reactions were observed. While the mechanical properties of the modified FRCs were suboptimal, the FRC intramedullary nails supported the femurs without structural failure, even in the cases of peri-implant fractures.<br /> (Copyright © 2014 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1878-0180
Volume :
40
Database :
MEDLINE
Journal :
Journal of the mechanical behavior of biomedical materials
Publication Type :
Academic Journal
Accession number :
25222871
Full Text :
https://doi.org/10.1016/j.jmbbm.2014.08.020