Dependence Independence Measure for Posterior and Anterior EMG Sensors Used in Simple and Complex Finger Flexion Movements: Evaluation Using SDICA.

Authors :
Naik GR
Baker KG
Nguyen HT
Source :
IEEE journal of biomedical and health informatics [IEEE J Biomed Health Inform] 2015 Sep; Vol. 19 (5), pp. 1689-1696. Date of Electronic Publication: 2014 Jul 17.
Publication Year :
2015

Abstract

Identification of simple and complex finger flexion movements using surface electromyography (sEMG) and a muscle activation strategy is necessary to control human-computer interfaces such as prosthesis and orthoses. In order to identify these movements, sEMG sensors are placed on both anterior and posterior muscle compartments of the forearm. In general, the accuracy of myoelectric classification depends on several factors, which include number of sensors, features extraction methods, and classification algorithms. Myoelectric classification using a minimum number of sensors and optimal electrode configuration is always a challenging task. Sometimes, using several sensors including high density electrodes will not guarantee high classification accuracy. In this research, we investigated the dependence and independence nature of anterior and posterior muscles during simple and complex finger flexion movements. The outcome of this research shows that posterior parts of the hand muscles are dependent and hence responsible for most of simple finger flexion. On the other hand, this study shows that anterior muscles are responsible for most complex finger flexion. This also indicates that simple finger flexion can be identified using sEMG sensors connected only on anterior muscles (making posterior placement either independent or redundant), and vice versa is true for complex actions which can be easily identified using sEMG sensors on posterior muscles. The result of this study is beneficial for optimal electrode configuration and design of prosthetics and other related devices using a minimum number of sensors.

Details

Language :
English
ISSN :
2168-2208
Volume :
19
Issue :
5
Database :
MEDLINE
Journal :
IEEE journal of biomedical and health informatics
Publication Type :
Academic Journal
Accession number :
25055388
Full Text :
https://doi.org/10.1109/JBHI.2014.2340397