Back to Search Start Over

Quantitative bioimaging of p-boronophenylalanine in thin liver tissue sections as a tool for treatment planning in boron neutron capture therapy.

Authors :
Reifschneider O
Schütz CL
Brochhausen C
Hampel G
Ross T
Sperling M
Karst U
Source :
Analytical and bioanalytical chemistry [Anal Bioanal Chem] 2015 Mar; Vol. 407 (9), pp. 2365-71.
Publication Year :
2015

Abstract

An analytical method using laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS) was developed and applied to assess enrichment of 10B-containing p-boronophenylalanine-fructose (BPA-f) and its pharmacokinetic distribution in human tissues after application for boron neutron capture therapy (BNCT). High spatial resolution (50 μm) and limits of detection in the low parts-per-billion range were achieved using a Nd:YAG laser of 213 nm wavelength. External calibration by means of 10B-enriched standards based on whole blood proved to yield precise quantification results. Using this calibration method, quantification of 10B in cancerous and healthy tissue was carried out. Additionally, the distribution of 11B was investigated, providing 10B enrichment in the investigated tissues. Quantitative imaging of 10B by means of LA-ICP-MS was demonstrated as a new option to characterise the efficacy of boron compounds for BNCT.

Details

Language :
English
ISSN :
1618-2650
Volume :
407
Issue :
9
Database :
MEDLINE
Journal :
Analytical and bioanalytical chemistry
Publication Type :
Academic Journal
Accession number :
25015045
Full Text :
https://doi.org/10.1007/s00216-014-8012-4