IUPAC-consistent approach to the limit of detection in partial least-squares calibration.

Authors :
Allegrini F
Olivieri AC
Source :
Analytical chemistry [Anal Chem] 2014 Aug 05; Vol. 86 (15), pp. 7858-66. Date of Electronic Publication: 2014 Jul 17.
Publication Year :
2014

Abstract

There is currently no well-defined procedure for providing the limit of detection (LOD) in multivariate calibration. Defining an estimator for the LOD in this scenario has shown to be more complex than intuitively extending the traditional univariate definition. For these reasons, although many attempts have been made to arrive at a reasonable convention, additional effort is required to achieve full agreement between the univariate and multivariate LOD definitions. In this work, a novel approach is presented to estimate the LOD in partial least-squares (PLS) calibration. Instead of a single LOD value, an interval of LODs is provided, which depends on the variation of the background composition in the calibration space. This is in contrast with previously proposed univariate extensions of the LOD concept. With the present definition, the LOD interval becomes a parameter characterizing the overall PLS calibration model, and not each test sample in particular, as has been proposed in the past. The new approach takes into account IUPAC official recommendations, and also the latest developments in error-in-variables theory for PLS calibration. Both simulated and real analytical systems have been studied for illustrating the properties of the new LOD concept.

Details

Language :
English
ISSN :
1520-6882
Volume :
86
Issue :
15
Database :
MEDLINE
Journal :
Analytical chemistry
Publication Type :
Academic Journal
Accession number :
25008998
Full Text :
https://doi.org/10.1021/ac501786u