Multicenter closed-loop insulin delivery study points to challenges for keeping blood glucose in a safe range by a control algorithm in adults and adolescents with type 1 diabetes from various sites.
- Source :
-
Diabetes technology & therapeutics [Diabetes Technol Ther] 2014 Oct; Vol. 16 (10), pp. 613-22. Date of Electronic Publication: 2014 Jul 08. - Publication Year :
- 2014
-
Abstract
- Background: The Control to Range Study was a multinational artificial pancreas study designed to assess the time spent in the hypo- and hyperglycemic ranges in adults and adolescents with type 1 diabetes while under closed-loop control. The controller attempted to keep the glucose ranges between 70 and 180 mg/dL. A set of prespecified metrics was used to measure safety.<br />Research Design and Methods: We studied 53 individuals for approximately 22 h each during clinical research center admissions. Plasma glucose level was measured every 15-30 min (YSI clinical laboratory analyzer instrument [YSI, Inc., Yellow Springs, OH]). During the admission, subjects received three mixed meals (1 g of carbohydrate/kg of body weight; 100 g maximum) with meal announcement and automated insulin dosing by the controller.<br />Results: For adults, the mean of subjects' mean glucose levels was 159 mg/dL, and mean percentage of values 71-180 mg/dL was 66% overall (59% daytime and 82% overnight). For adolescents, the mean of subjects' mean glucose levels was 166 mg/dL, and mean percentage of values in range was 62% overall (53% daytime and 82% overnight). Whereas prespecified criteria for safety were satisfied by both groups, they were met at the individual level in adults only for combined daytime/nighttime and for isolated nighttime. Two adults and six adolescents failed to meet the daytime criterion, largely because of postmeal hyperglycemia, and another adolescent failed to meet the nighttime criterion.<br />Conclusions: The control-to-range system performed as expected: faring better overnight than during the day and performing with variability between patients even after individualization based on patients' prior settings. The system had difficulty preventing postmeal excursions above target range.
- Subjects :
- Adolescent
Adult
Algorithms
Blood Glucose Self-Monitoring
Diabetes Mellitus, Type 1 drug therapy
Dietary Carbohydrates
Female
Humans
Hyperglycemia blood
Hypoglycemia blood
Insulin metabolism
Insulin Secretion
Male
Meals
Monitoring, Physiologic
Patient Safety
Pilot Projects
Postprandial Period
Reproducibility of Results
Time Factors
Treatment Outcome
Blood Glucose metabolism
Diabetes Mellitus, Type 1 blood
Glycated Hemoglobin metabolism
Hyperglycemia prevention & control
Hypoglycemia prevention & control
Hypoglycemic Agents administration & dosage
Insulin administration & dosage
Pancreas, Artificial
Subjects
Details
- Language :
- English
- ISSN :
- 1557-8593
- Volume :
- 16
- Issue :
- 10
- Database :
- MEDLINE
- Journal :
- Diabetes technology & therapeutics
- Publication Type :
- Academic Journal
- Accession number :
- 25003311
- Full Text :
- https://doi.org/10.1089/dia.2014.0066