Back to Search Start Over

Development and performance evaluation of an electromagnetic-type shock wave generator for lipolysis.

Authors :
Liang SM
Chang MH
Yang ZY
Source :
The Review of scientific instruments [Rev Sci Instrum] 2014 Jan; Vol. 85 (1), pp. 015113.
Publication Year :
2014

Abstract

This study aims at the design and development of electromagnetic-type intermittent shock wave generation in a liquid. The shock wave generated is focused at a focal point through an acoustic lens. This hardware device mainly consists of a full-wave bridge rectifier, 6 capacitors, a spark gap, and a flat coil. A metal disk is mounted in a liquid-filled tube and is placed in close proximity to the flat coil. Due to the repulsive force existing between the coil and disk shock waves are generated, while an eddy current is induced in the metal disk. Some components and materials associated with the device are also described. By increasing the capacitance content to enhance electric energy level, a highly focused pressure can be achieved at the focal point through an acoustic lens in order to lyse fat tissue. Focused pressures were measured at the focal point and its vicinity for different operation voltages. The designed shock wave generator with an energy intensity of 0.0016 mJ/mm(2) (at 4 kV) and 2000 firings or higher energy intensities with 1000 firings is found to be able to disrupt pig fat tissue.

Details

Language :
English
ISSN :
1089-7623
Volume :
85
Issue :
1
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
24517818
Full Text :
https://doi.org/10.1063/1.4861178