Abstract

A rapid and simple validated experimental protocol is proposed for the accurate determination of the true intrinsic column efficiency and for that of the variance of the extra-column volume of the instrument used, the latter being obtained without requiring the removal of the chromatographic column from the HPLC system. This protocol was applied to 2.1mm×100mm columns packed with sub-3 (2.7μm Halo Peptide ES-C18) and sub-2μm (1.6μm prototype) core-shell particles. It was validated by observing the linear behavior of the plot of the apparent column plate height versus the reciprocal of (1+k')(2) for at least three homologous compounds, with a linear regression coefficient R(2) larger than 0.999. Irrespective of the contribution of the several, different instruments used to the total band broadening, the same column HETP value was obtained within 5%. This new protocol outperform the classical one in which the chromatographic column is replaced with a zero dead volume (ZDV) union connector to measure the extra-column volume variance, which is subtracted from the variance measured with the column to measure the intrinsic HETP. This protocol fails because it significantly underestimates the system volume variance.<br /> (Copyright © 2013 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-3778
Volume :
1327
Database :
MEDLINE
Journal :
Journal of chromatography. A
Publication Type :
Academic Journal
Accession number :
24438831
Full Text :
https://doi.org/10.1016/j.chroma.2013.12.003