Back to Search
Start Over
Beta spectrometry with metallic magnetic calorimeters.
- Source :
-
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2014 May; Vol. 87, pp. 302-5. Date of Electronic Publication: 2013 Nov 28. - Publication Year :
- 2014
-
Abstract
- Metallic magnetic calorimeters are a specific type of cryogenic detectors that have been shown to enable precise measurement of the shape of low energy beta spectra. The aim of their use at LNHB is the determination of the shape factors of beta spectra. The beta source is enclosed in the detector absorber, allowing for very high detection efficiency. It has turned out that the type of source is of crucial importance for the correctness of the measured spectrum. Spectra of (63)Ni measured with several sources prepared by drying a NiCl2 solution differ from one another and from theory, whereas spectra measured with electroplated sources are reproducible and agree with theory. With these latter measurements we could confirm the atomic exchange effect down to very low energy (200 eV).<br /> (© 2013 Published by Elsevier Ltd.)
Details
- Language :
- English
- ISSN :
- 1872-9800
- Volume :
- 87
- Database :
- MEDLINE
- Journal :
- Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- Publication Type :
- Academic Journal
- Accession number :
- 24368065
- Full Text :
- https://doi.org/10.1016/j.apradiso.2013.11.024