Back to Search Start Over

Determination of dead-layer variation in HPGe detectors.

Authors :
Andreotti E
Hult M
Marissens G
Lutter G
Garfagnini A
Hemmer S
von Sturm K
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2014 May; Vol. 87, pp. 331-5. Date of Electronic Publication: 2013 Nov 26.
Publication Year :
2014

Abstract

The dead-layer uniformity of the top surface of two high purity germanium detectors has been studied using a novel automated scanning set-up that allows a fine-grained topography of a detector's top and lateral surfaces. Comparisons between measurements and Monte Carlo simulations allowed implementation of a dead-layer variation into the detector model, which reproduces the measurements results. The effect of the non-uniform dead-layer on activity determinations based on low-energy γ-rays (i.e. below ~100 keV) has been determined to be of the order of 10% or more.<br /> (© 2013 Published by Elsevier Ltd.)