Back to Search Start Over

Evaluation of a personal and environmental dosemeter based on CR-39 track detectors in quasi-monoenergetic neutron fields.

Authors :
Caresana M
Ferrarini M
Parravicini A
Sashala Naik A
Source :
Radiation protection dosimetry [Radiat Prot Dosimetry] 2014 Oct; Vol. 161 (1-4), pp. 100-3. Date of Electronic Publication: 2013 Dec 08.
Publication Year :
2014

Abstract

In this paper, the evaluation of the dosimetric capability of a detector based on a CR-39 solid-state nuclear track detector coupled to a 1 cm thickness of PMMA radiator was made with the aim of understanding the applicability of this technique to personal and environmental neutron dosimetry. The dosemeter has been exposed to monoenergetic and quasi-monoenergetic neutron beams at PTB in Braunschweig, Germany and at Ithemba Laboratories, in Faure, South Africa, with peak energies ranging from 0.565 to 100 MeV. The results showed a response that is almost independent of the neutron energy in the whole energy range.<br /> (© The Author 2013. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1742-3406
Volume :
161
Issue :
1-4
Database :
MEDLINE
Journal :
Radiation protection dosimetry
Publication Type :
Academic Journal
Accession number :
24324248
Full Text :
https://doi.org/10.1093/rpd/nct320