Field evaluation of a prototype paper-based point-of-care fingerstick transaminase test.

Authors :
Pollock NR
McGray S
Colby DJ
Noubary F
Nguyen H
Nguyen TA
Khormaee S
Jain S
Hawkins K
Kumar S
Rolland JP
Beattie PD
Chau NV
Quang VM
Barfield C
Tietje K
Steele M
Weigl BH
Source :
PloS one [PLoS One] 2013 Sep 30; Vol. 8 (9), pp. e75616. Date of Electronic Publication: 2013 Sep 30 (Print Publication: 2013).
Publication Year :
2013

Abstract

Monitoring for drug-induced liver injury (DILI) via serial transaminase measurements in patients on potentially hepatotoxic medications (e.g., for HIV and tuberculosis) is routine in resource-rich nations, but often unavailable in resource-limited settings. Towards enabling universal access to affordable point-of-care (POC) screening for DILI, we have performed the first field evaluation of a paper-based, microfluidic fingerstick test for rapid, semi-quantitative, visual measurement of blood alanine aminotransferase (ALT). Our objectives were to assess operational feasibility, inter-operator variability, lot variability, device failure rate, and accuracy, to inform device modification for further field testing. The paper-based ALT test was performed at POC on fingerstick samples from 600 outpatients receiving HIV treatment in Vietnam. Results, read independently by two clinic nurses, were compared with gold-standard automated (Roche Cobas) results from venipuncture samples obtained in parallel. Two device lots were used sequentially. We demonstrated high inter-operator agreement, with 96.3% (95% C.I., 94.3-97.7%) agreement in placing visual results into clinically-defined "bins" (<3x, 3-5x, and >5x upper limit of normal), >90% agreement in validity determination, and intraclass correlation coefficient of 0.89 (95% C.I., 0.87-0.91). Lot variability was observed in % invalids due to hemolysis (21.1% for Lot 1, 1.6% for Lot 2) and correlated with lots of incorporated plasma separation membranes. Invalid rates <1% were observed for all other device controls. Overall bin placement accuracy for the two readers was 84% (84.3%/83.6%). Our findings of extremely high inter-operator agreement for visual reading-obtained in a target clinical environment, as performed by local practitioners-indicate that the device operation and reading process is feasible and reproducible. Bin placement accuracy and lot-to-lot variability data identified specific targets for device optimization and material quality control. This is the first field study performed with a patterned paper-based microfluidic device and opens the door to development of similar assays for other important analytes.

Details

Language :
English
ISSN :
1932-6203
Volume :
8
Issue :
9
Database :
MEDLINE
Journal :
PloS one
Publication Type :
Academic Journal
Accession number :
24098705
Full Text :
https://doi.org/10.1371/journal.pone.0075616