Back to Search Start Over

Design and initial evaluation of a treatment planning software system for MRI-guided laser ablation in the brain.

Authors :
Yeniaras E
Fuentes DT
Fahrenholtz SJ
Weinberg JS
Maier F
Hazle JD
Stafford RJ
Source :
International journal of computer assisted radiology and surgery [Int J Comput Assist Radiol Surg] 2014 Jul; Vol. 9 (4), pp. 659-67. Date of Electronic Publication: 2013 Oct 05.
Publication Year :
2014

Abstract

Purpose: An open-source software system for planning magnetic resonance (MR)-guided laser-induced thermal therapy (MRgLITT) in brain is presented. The system was designed to provide a streamlined and operator-friendly graphical user interface (GUI) for simulating and visualizing potential outcomes of various treatment scenarios to aid in decisions on treatment approach or feasibility.<br />Methods: A portable software module was developed on the 3D Slicer platform, an open-source medical imaging and visualization framework. The module introduces an interactive GUI for investigating different laser positions and power settings as well as the influence of patient-specific tissue properties for quickly creating and evaluating custom treatment options. It also provides a common treatment planning interface for use by both open-source and commercial finite element solvers. In this study, an open-source finite element solver for Pennes' bioheat equation is interfaced to the module to provide rapid 3D estimates of the steady-state temperature distribution and potential tissue damage in the presence of patient-specific tissue boundary conditions identified on segmented MR images.<br />Results: The total time to initialize and simulate an MRgLITT procedure using the GUI was [Formula: see text]5 min. Each independent simulation took [Formula: see text]30 s, including the time to visualize the results fused with the planning MRI. For demonstration purposes, a simulated steady-state isotherm contour [Formula: see text] was correlated with MR temperature imaging (N = 5). The mean Hausdorff distance between simulated and actual contours was 2.0 mm [Formula: see text], whereas the mean Dice similarity coefficient was 0.93 [Formula: see text].<br />Conclusions: We have designed, implemented, and conducted initial feasibility evaluations of a software tool for intuitive and rapid planning of MRgLITT in brain. The retrospective in vivo dataset presented herein illustrates the feasibility and potential of incorporating fast, image-based bioheat predictions into an interactive virtual planning environment for such procedures.

Details

Language :
English
ISSN :
1861-6429
Volume :
9
Issue :
4
Database :
MEDLINE
Journal :
International journal of computer assisted radiology and surgery
Publication Type :
Academic Journal
Accession number :
24091853
Full Text :
https://doi.org/10.1007/s11548-013-0948-x