Long-term outcome of a metal-on-polyethylene cementless hip resurfacing.
- Source :
-
The Journal of arthroplasty [J Arthroplasty] 2014 Apr; Vol. 29 (4), pp. 802-7. Date of Electronic Publication: 2013 Oct 03. - Publication Year :
- 2014
-
Abstract
- Due to the well-documented problems surrounding metal-on-metal bearings, the use of hip resurfacing has declined. Since the potential benefits of hip resurfacing remain desirable, it may be beneficial to investigate the long-term outcome of hip resurfacings using metal-on-polyethylene in the 1980's. We report the long-term survivorship and modes of failure of a cementless metal-on-polyethylene resurfacing (n = 178) with different porous ingrowth surfaces. While acetabular loosening was absent, a high incidence of femoral failures (femoral loosening = 18.1%, osteolytic neck fracture = 21%) occurred despite using the same ingrowth surface for both components. Ongoing developments using the lessons learned from these previous generation components and utilizing modern low wear materials, e.g., cross-linked polyethylene, may lead to improved implants for future hip resurfacings.<br /> (© 2014 Elsevier Inc. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1532-8406
- Volume :
- 29
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- The Journal of arthroplasty
- Publication Type :
- Academic Journal
- Accession number :
- 24090660
- Full Text :
- https://doi.org/10.1016/j.arth.2013.08.016