Digital arthrodesis with a one-piece memory nitinol intramedullary fixation device: a retrospective review.
- Source :
-
Foot & ankle specialist [Foot Ankle Spec] 2013 Oct; Vol. 6 (5), pp. 364-6. - Publication Year :
- 2013
-
Abstract
- Unlabelled: A 1-piece memory Nitinol intramedullary fixation device (Smart Toe; Stryker Corporation, Kalamazoo, MI) was used for proximal interphalangeal joint arthrodesis for correction of painful hammertoes in digits 2, 3, and 4. Sixty-five implants were placed in 35 patients. The mean age of our patients was 62.2 years (range = 27-82; standard deviation = 12.5). Mean follow-up time was 27 months (range = 12-40 months; standard deviation = 7). Overall, a 93.8% fusion rate was noted. Complications were noted in 4 patients (6.1%): 1 asymptomatic nonunion (1.5%), 2 hardware failures (3%), and 1 implant displacement (1.5%). To date, no patients required revisional surgery or hardware removal.<br />Level of Evidence: V.
- Subjects :
- Adult
Aged
Aged, 80 and over
Alloys therapeutic use
Arthrodesis methods
Female
Hammer Toe Syndrome diagnostic imaging
Humans
Male
Middle Aged
Radiography
Retrospective Studies
Toe Joint diagnostic imaging
Treatment Outcome
Arthrodesis instrumentation
Hammer Toe Syndrome surgery
Internal Fixators
Toe Joint surgery
Subjects
Details
- Language :
- English
- ISSN :
- 1938-7636
- Volume :
- 6
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Foot & ankle specialist
- Publication Type :
- Academic Journal
- Accession number :
- 24026289
- Full Text :
- https://doi.org/10.1177/1938640013496458