Back to Search
Start Over
Implantable cardioverter-defibrillators in hypertrophic cardiomyopathy: patient outcomes, rate of appropriate and inappropriate interventions, and complications.
- Source :
-
American heart journal [Am Heart J] 2013 Sep; Vol. 166 (3), pp. 496-502. Date of Electronic Publication: 2013 Jul 16. - Publication Year :
- 2013
-
Abstract
- Background: Sudden cardiac death (SCD) is the most devastating complication of hypertrophic cardiomyopathy (HCM), but this can be prevented by an implantable cardioverter-defibrillator (ICD). The aim of this study is to evaluate HCM patients with ICDs for primary or secondary prevention of SCD.<br />Methods: The study population consisted of all HCM patients with an ICD in 2 tertiary referral clinics. End points during follow-up were total and cardiac mortality, appropriate and inappropriate ICD intervention, and device-related complications. Cox-regression analysis was performed to identify predictors of outcome.<br />Results: ICDs were implanted in 134 patients with HCM (mean age 44 ± 17 years, 34% women, 4.2 ± 4.8 years follow-up). Annualized cardiac mortality rate was 3.4% per year and associated with New York Heart Association class III or IV (HR 5.2 [2.0-14, P = .002]) and cardiac resynchronization therapy (HR 6.3 [2.1-20, P = .02]). Appropriate ICD interventions occurred in 38 patients (6.8%/year) and was associated with implantation for secondary prevention of SCD (HR 4.0 [1.8-9.1], P = .001) and male gender (HR 3.3 [1.2-9.0], P = .02). Inappropriate ICD intervention occurred in 21 patients (3.7%/year) and in 20 patients device related complications were documented (3.6%/year).<br />Conclusion: ICDs successfully abort life-threatening arrhythmias in HCM patients at increased risk of SCD with an annualized intervention rate of 6.8% per year. End-stage heart failure is the main cause of mortality in these patients. The annualized rate of inappropriate ICD intervention was 3.7% per year, whereas device-related complications occurred 3.6% per year.<br /> (© 2013.)
- Subjects :
- Adult
Cardiomyopathy, Hypertrophic mortality
Death, Sudden, Cardiac prevention & control
Female
Follow-Up Studies
Humans
Male
Middle Aged
Regression Analysis
Risk Assessment
Risk Factors
Treatment Outcome
Cardiac Resynchronization Therapy statistics & numerical data
Cardiomyopathy, Hypertrophic therapy
Death, Sudden, Cardiac etiology
Defibrillators, Implantable adverse effects
Subjects
Details
- Language :
- English
- ISSN :
- 1097-6744
- Volume :
- 166
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- American heart journal
- Publication Type :
- Academic Journal
- Accession number :
- 24016499
- Full Text :
- https://doi.org/10.1016/j.ahj.2013.06.009