The cryo-needle: a new tool for histological biopsies. A feasibility study.
- Source :
-
Lung [Lung] 2013 Dec; Vol. 191 (6), pp. 611-7. Date of Electronic Publication: 2013 Aug 30. - Publication Year :
- 2013
-
Abstract
- Purpose: Endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is a standard procedure for intrathoracic lymph node biopsies. The newly developed cryo-needle operates in a similar way to the EBUS-TBNA but is able to obtain specimens for histological evaluation. The purpose of this animal study was to evaluate the feasibility, effect, and safety of the cryo-needle biopsies.<br />Methods: Four EBUS-guided cryo-needle biopsies were obtained from a mediastinal lymph node of a healthy pig. In an open surgery approach, cryo-needle biopsies using activation times of 1, 2, and 3 s (A1/A2/A3) and needle biopsies using a 21-gauge EBUS-TBNA needle were obtained from mesenteric lymph nodes. Cryo-needle biopsies A2 were performed with (A2+) and without (A2-) an oversheath. The size, weight, percentage of lymphatic tissue and artefact-free area of each cryobiopsy were evaluated. Smears were made with the TBNA-needle aspirates to determine the number of lymphocytes per high-power field (HPF). The bleeding duration was measured.<br />Results: We successfully obtained EBUS-guided cryo-needle biopsies. The area and weight of the biopsies A3 and A2+ were significantly larger compared with A1 (1.7 ± 0.8 and 1.4 ± 0.3 vs. 0.9 ± 0.4 mm(2); 5.2 ± 2.4 and 3.4 ± 1.8 vs. 1.5 ± 0.7 mg). The percentage of lymphatic tissue of the cryobiopsies was 90 ± 25 and 98 % of samples were artefact-free. The number of lymphocytes/HPF of TBNA-needle smears was 128 ± 54.3. There was no difference in bleeding duration between the techniques.<br />Conclusions: The cryo-needle yields large histological specimens of high quality.
- Subjects :
- Animals
Artifacts
Endoscopic Ultrasound-Guided Fine Needle Aspiration adverse effects
Equipment Design
Feasibility Studies
Female
Hemorrhage etiology
Lymphocyte Count
Models, Animal
Predictive Value of Tests
Swine
Time Factors
Endoscopic Ultrasound-Guided Fine Needle Aspiration instrumentation
Freezing
Lymph Nodes pathology
Lymphocytes pathology
Needles
Subjects
Details
- Language :
- English
- ISSN :
- 1432-1750
- Volume :
- 191
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Lung
- Publication Type :
- Academic Journal
- Accession number :
- 23990134
- Full Text :
- https://doi.org/10.1007/s00408-013-9502-4