Back to Search Start Over

Filling the void: a coalescent numerical and experimental technique to determine aortic stent graft mechanics.

Authors :
De Bock S
Iannaccone F
De Beule M
Van Loo D
Vermassen F
Verhegghe B
Segers P
Source :
Journal of biomechanics [J Biomech] 2013 Sep 27; Vol. 46 (14), pp. 2477-82. Date of Electronic Publication: 2013 Jul 27.
Publication Year :
2013

Abstract

The presented study details a combined experimental and computational method to assess and compare the mechanical behavior of the main body of 4 different stent graft designs. The mechanical response to a flat plate compression and radial crimping of the devices is derived and related to geometrical and material features of different stent designs. The finite element modeling procedure is used to complement the experimental results and conduct a solution sensitivity study. Finite element evaluations of the mechanical behavior match well with experimental findings and are used as a quantitative basis to discuss design characteristics of the different devices.<br /> (Copyright © 2013 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1873-2380
Volume :
46
Issue :
14
Database :
MEDLINE
Journal :
Journal of biomechanics
Publication Type :
Academic Journal
Accession number :
23953501
Full Text :
https://doi.org/10.1016/j.jbiomech.2013.07.010