Back to Search
Start Over
Study to determine the repeatability of supra-sternal Doppler (ultrasound cardiac output monitor) during general anaesthesia: effects of scan quality, flow volume, and increasing age.
- Source :
-
British journal of anaesthesia [Br J Anaesth] 2013 Dec; Vol. 111 (6), pp. 907-15. Date of Electronic Publication: 2013 Jul 23. - Publication Year :
- 2013
-
Abstract
- Background: The ultrasound cardiac output monitor (USCOM) is a continuous wave Doppler system designed to measure cardiac output (CO) non-invasively and intermittently either from the pulmonary or from the aortic valve. USCOM scan quality is critical to obtaining reliable data and during anaesthesia it is said to deteriorate with increasing age. The aim of this study was to investigate the effect of age on supra-sternal USCOM scan repeatability during anaesthesia.<br />Methods: We performed a series of 6 USCOM scans in 180 patients of all ages after induction for routine surgery. A 12-point Cattermole (CS) score and 10-point insonation (IS) score were used to evaluate scan quality and ease of insonation. The coefficients of variation (CVs) of USCOM variables [CO, peak velocity, stroke volume index (SVI) and the corrected flow time] were derived from the series of six readings.<br />Results: In >95% of young patients (age <50 yr), it was easy to obtain a good-quality USCOM scan (CS>8). In these patients, repeatability of serial readings was good with CVs<5% and precision of less than ±10%. In older patients (>50 yr), scan quality and ease of insonation declined, with >25% of patients >60 yr having unreliable USCOM scans (CS<5). In these patients, the CV was >5-10%. In several elderly patients (>65 yr), we failed to locate the USCOM signal. Average scan time increased with age (30 to >60 s). SVI was also strongly correlated with scan quality (R(2)=0.77).<br />Conclusions: Increasing age has a significant effect on USCOM scan quality and data reliability.
- Subjects :
- Adolescent
Adult
Age Factors
Aged
Aortic Valve diagnostic imaging
Aortic Valve physiopathology
Blood Flow Velocity physiology
Child
Echocardiography, Doppler methods
Echocardiography, Doppler standards
Female
Humans
Male
Middle Aged
Monitoring, Intraoperative standards
Reproducibility of Results
Sternum
Aging physiology
Anesthesia, General
Cardiac Output physiology
Monitoring, Intraoperative methods
Subjects
Details
- Language :
- English
- ISSN :
- 1471-6771
- Volume :
- 111
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- British journal of anaesthesia
- Publication Type :
- Academic Journal
- Accession number :
- 23880100
- Full Text :
- https://doi.org/10.1093/bja/aet254