Back to Search
Start Over
Photoinduced oxidation of a tris(2,2'-bipyridyl)ruthenium(II)-peroxodisulfate chemiluminescence system for the analysis of mebeverine HCl pharmaceutical formulations and biological fluids using a two-chip device.
- Source :
-
Luminescence : the journal of biological and chemical luminescence [Luminescence] 2014 May; Vol. 29 (3), pp. 275-83. Date of Electronic Publication: 2013 Jun 13. - Publication Year :
- 2014
-
Abstract
- A new method for the analysis of mebeverine hydrochloride (MEB) has been developed using a two-chip device. The method is highly selective, sensitive, rapid and consumes minute amount of reagents. The developed method is free of interference from the degradation products of MEB and from common ingredients present in pharmaceutical formulations. The limit of detection was 0.043 µg/mL, and the limit of quantification was 0.138 µg/mL. The short analysis time per sample (20 s) allowed a large number of analyses to be performed within a very short time. Various samples were analyzed, including two different pharmaceutical formulations and a uniformity of content analysis for 20 tablets from a known batch and two biological samples at different concentrations. In addition, the method was compared with a validated high-performance liquid chromatography (HPLC) method and the results clearly indicated the suitability of the developed method for routine analyses. A new mechanism for the tris(2,2'-bipyridyl)ruthenium(II) (Ru(bpy)3 (2+) )-peroxodisulfate (S2 O8 (2-) ) chemiluminescence (CL) system has also been proposed. The mechanism is based on photoinduced oxidation of Ru(bpy)3 (2+) to Ru(bpy)3 (3+) via the formation of Ru(bpy)3 (2+) * upon irradiation with visible light. S2 O8 (2-) then oxidizes Ru(bpy)3 (2+) * to Ru(bpy)3 (3+) and the analyte subsequently reduces the resultant Ru(bpy)3 (3+) to Ru(bpy)3 (2+) *, which then produces the CL signal.<br /> (Copyright © 2013 John Wiley & Sons, Ltd.)
- Subjects :
- Humans
Lab-On-A-Chip Devices
Light
Luminescent Measurements instrumentation
Oxidation-Reduction radiation effects
Phenethylamines urine
Photochemical Processes
Tablets analysis
Luminescent Measurements methods
Organometallic Compounds chemistry
Pharmaceutical Preparations analysis
Phenethylamines analysis
Sulfates chemistry
Subjects
Details
- Language :
- English
- ISSN :
- 1522-7243
- Volume :
- 29
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Luminescence : the journal of biological and chemical luminescence
- Publication Type :
- Academic Journal
- Accession number :
- 23760778
- Full Text :