Manufacturing lot affects polyethylene tibial insert volume, thickness, and surface geometry.

Authors :
Teeter MG
Milner JS
MacDonald SJ
Naudie DD
Source :
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine [Proc Inst Mech Eng H] 2013 Aug; Vol. 227 (8), pp. 884-9. Date of Electronic Publication: 2013 Apr 26.
Publication Year :
2013

Abstract

To perform wear measurements on retrieved joint replacement implants, a reference geometry of the implant's original state is required. Since implants are rarely individually scanned before implantation, a different, new implant of the same kind and size is frequently used. However, due to manufacturing variability, errors may be introduced into these measurements, as the dimensions between the retrieved and reference components may not be exactly the same. The hypothesis of this study was that new polyethylene tibial inserts from different manufacturing lots would demonstrate greater variability than those from the same lot. In total, 12 new tibial inserts of the same model and size were obtained, 5 from the same lot and the remainder from different lots. The geometry of each tibial insert was obtained using microcomputed tomography. Measurements of tibial insert volume, thickness, and three-dimensional surface deviations were obtained and compared between tibial inserts from the same and different manufacturing lots. Greater variability was found for the tibial inserts from different manufacturing lots for all types of measurements, including a fourfold difference in volume variability (p < 0.001) and a maximum of 0.21 mm difference in thickness (p < 0.001). Investigators should be aware of this potential confounding error and take steps to minimize it, such as by averaging together the geometries of multiple new components from different manufacturing lots for use as the reference geometry.

Details

Language :
English
ISSN :
2041-3033
Volume :
227
Issue :
8
Database :
MEDLINE
Journal :
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
Publication Type :
Academic Journal
Accession number :
23636769
Full Text :
https://doi.org/10.1177/0954411913486755