Back to Search Start Over

Preoperative analysis of the stability of fit of a patient-specific surgical guide.

Authors :
Van den Broeck J
Wirix-Speetjens R
Vander Sloten J
Source :
Computer methods in biomechanics and biomedical engineering [Comput Methods Biomech Biomed Engin] 2015; Vol. 18 (1), pp. 38-47. Date of Electronic Publication: 2013 Apr 29.
Publication Year :
2015

Abstract

Although the use of patient-specific surgical guides has gained popularity over the past decade, little research has been done to examine in an objective and qualitative way the fit of such instruments. In this study, we have developed a model to predict the stability of a guide designed to fit on a supporting bone surface, thereby providing feedback on the translational and rotational stability of the device. The method was validated by comparing different guide designs with respect to their stability on the contact surface and comparing these results to those measured with a set of experiments. This validation experiment indicates that our stability model can be used to predict the stability of the fit of a surgical guide during the preoperative design process.

Details

Language :
English
ISSN :
1476-8259
Volume :
18
Issue :
1
Database :
MEDLINE
Journal :
Computer methods in biomechanics and biomedical engineering
Publication Type :
Academic Journal
Accession number :
23627973
Full Text :
https://doi.org/10.1080/10255842.2013.774383