Back to Search Start Over

The effective temperature of ions stored in a linear quadrupole ion trap mass spectrometer.

Authors :
Donald WA
Khairallah GN
O'Hair RA
Source :
Journal of the American Society for Mass Spectrometry [J Am Soc Mass Spectrom] 2013 Jun; Vol. 24 (6), pp. 811-5. Date of Electronic Publication: 2013 Apr 20.
Publication Year :
2013

Abstract

The extent of internal energy deposition into ions upon storage, radial ejection, and detection using a linear quadrupole ion trap mass spectrometer is investigated as a function of ion size (m/z 59 to 810) using seven ion-molecule thermometer reactions that have well characterized reaction entropies and enthalpies. The average effective temperatures of the reactants and products of the ion-molecule reactions, which were obtained from ion-molecule equilibrium measurements, range from 295 to 350 K and do not depend significantly on the number of trapped ions, m/z value, ion trap q z value, reaction enthalpy/entropy, or the number of vibrational degrees of freedom for the seven reactions investigated. The average of the effective temperature values obtained for all seven thermometer reactions is 318 ± 23 K, which indicates that linear quadrupole ion trap mass spectrometers can be used to study the structure(s) and reactivity of ions at near ambient temperature.

Details

Language :
English
ISSN :
1879-1123
Volume :
24
Issue :
6
Database :
MEDLINE
Journal :
Journal of the American Society for Mass Spectrometry
Publication Type :
Academic Journal
Accession number :
23605688
Full Text :