Intratester test-retest reliability of insertion gain measures.
- Source :
-
Ear and hearing [Ear Hear] 1990 Jun; Vol. 11 (3), pp. 181-4. - Publication Year :
- 1990
-
Abstract
- Intratester test-retest reliability of insertion gain was determined on 28 subjects using the Frye 6500 real-ear analyzer. Results revealed mean differences of less than 1 dB for repeat measurements at six test frequencies between 250 and 4000 Hz. Also, over 80% of the repeat measures were within +/- 3 dB of the first measure. The results of this study are compared with previous studies of intratester reliability of insertion gain measures.
Details
- Language :
- English
- ISSN :
- 0196-0202
- Volume :
- 11
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Ear and hearing
- Publication Type :
- Academic Journal
- Accession number :
- 2358127
- Full Text :
- https://doi.org/10.1097/00003446-199006000-00003