Back to Search Start Over

Bolus timing in high-pitch CT angiography of the aorta.

Authors :
Beeres M
Loch M
Schulz B
Kerl M
Al-Butmeh F
Bodelle B
Herrmann E
Gruber-Rouh T
Lee C
Jacobi V
Vogl TJ
Bauer RW
Source :
European journal of radiology [Eur J Radiol] 2013 Jun; Vol. 82 (6), pp. 1028-33. Date of Electronic Publication: 2013 Jan 31.
Publication Year :
2013

Abstract

Objective: To investigate the bolus geometry in high-pitch CT angiography (CTA) of the aorta without ECG synchronisation in comparison to single-source CT.<br />Methods: Overall 160 consecutive patients underwent CTA either in conventional single-source mode with a pitch of 1.2 (group 1), or in dual-source mode with a pitch of 3.0 (groups 2, 3 and 4) using different contrast media timings with bolus triggering at 140 HU (5s, group 1; 10s, group 2; 12s, group 3; 14s, group 4). Contrast material, saline flush, flow rate and kV/mAs settings were kept equal for optimum comparability. Aortic attenuation was measured along the z-axis of the patient at different anatomic landmarks and subjective image quality was compared.<br />Results: The most homogeneous enhancement of the aorta was reached with a delay of 10s after reaching the trigger threshold. The imaging length was not significantly different, but the examination time was significantly (p<0.001) shorter in the high-pitch group (7.7s vs. 1.7s for group 1 vs. 2, 3 and 4).<br />Conclusion: In high-pitch CT angiography using a start delay of 10s after a trigger threshold of 140 HU in the descending aorta is reached, a homogenous contrast along the z-axis is accomplished.<br /> (Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1872-7727
Volume :
82
Issue :
6
Database :
MEDLINE
Journal :
European journal of radiology
Publication Type :
Academic Journal
Accession number :
23375815
Full Text :
https://doi.org/10.1016/j.ejrad.2013.01.004