Evaluation of an algorithmic approach in comparison with the Illumigene assay for laboratory diagnosis of Clostridium difficile infection.

Authors :
Walkty A
Lagacé-Wiens PR
Manickam K
Adam H
Pieroni P
Hoban D
Karlowsky JA
Alfa M
Source :
Journal of clinical microbiology [J Clin Microbiol] 2013 Apr; Vol. 51 (4), pp. 1152-7. Date of Electronic Publication: 2013 Jan 30.
Publication Year :
2013

Abstract

The following three diagnostic algorithms were evaluated in comparison with the Illumigene assay as a stand-alone test for Clostridium difficile detection: glutamate dehydrogenase antigen screen (GDH) followed by toxin A/B antigen testing (Tox A/B) with the cell cytotoxicity assay for discordant specimens (algorithm 1), GDH followed by the Illumigene (algorithm 2), and GDH followed by Tox A/B with the Illumigene for discordant specimens (algorithm 3). A total of 428 stool specimens submitted to three clinical microbiology laboratories in Manitoba, Canada, for C. difficile detection between June 2011 and April 2012 were included in the study. The prevalence of C. difficile in the stool specimens was 14.7% (63/428) based on toxigenic culture (microbiologic reference standard). The sensitivity and specificity of the Illumigene for C. difficile detection were 73.0% and 99.7%, respectively. The corresponding sensitivities and specificities were 65.1% and 100.0% for algorithm 1, 68.3% and 100.0% for algorithm 2, and 69.8% and 100.0% for algorithm 3. Using algorithm 1, a cell cytotoxicity assay was required for toxin detection in 37% of positive tests, prolonging turnaround time. However, the predictive value of a positive test based on a clinical reference standard (all tests positive or cytotoxigenic culture positive and clinical disease on chart review) was slightly higher with algorithm 1 than with the Illumigene assay as a stand-alone test or as part of an algorithm (algorithms 2 and 3). Based on a reduction in turnaround time, simplicity, and acceptable sensitivity and specificity, we recommend algorithm 2 (screening with the GDH antigen test and confirmatory testing with the Illumigene).

Details

Language :
English
ISSN :
1098-660X
Volume :
51
Issue :
4
Database :
MEDLINE
Journal :
Journal of clinical microbiology
Publication Type :
Academic Journal
Accession number :
23363829
Full Text :
https://doi.org/10.1128/JCM.03203-12