Fast and sensitive UHPLC methods with fluorescence and tandem mass spectrometry detection for the determination of tetracycline antibiotics in surface waters.

Authors :
Škrášková K
Santos LH
Šatínský D
Pena A
Montenegro MC
Solich P
Nováková L
Source :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences [J Chromatogr B Analyt Technol Biomed Life Sci] 2013 May 15; Vol. 927, pp. 201-8. Date of Electronic Publication: 2013 Jan 09.
Publication Year :
2013

Abstract

In this paper two fast and highly sensitive ultra-high performance liquid chromatography (UHPLC) methods for the determination of tetracycline antibiotics (oxytetracycline, tetracycline, doxycycline, demeclocycline, chlortetracycline, minocycline and degradation product epitetracycline) in surface waters have been developed using fluorescence (FL) and mass spectrometry (MS) detection. ACQUITY UPLC BEH C8 and ACQUITY CSH C18 columns were employed for FL and MS detection, respectively, both packed with 1.7μm particles. Mixed-mode separation mechanism of CSH (charged surface technology) sorbent was found particularly useful in analysis of TCs, which possess problematic amphoteric structures. The FL methodology was based on chelation of tetracyclines with calcium ions to perform on-column derivatisation. The developed methods were compared in the terms of validation parameters including linearity, sensitivity, precision and accuracy. The linearity range for FL detection was within 7ngmL(-1) to 50μgmL(-1) with method limit of detection (MLOD) as low as 0.2ngmL(-1) for most of the analytes. MS detection showed even higher sensitivity reaching MLOD of 0.003ngmL(-1), which is the highest sensitivity reported so far in analysis of TCs. Matrix matched calibration curves in the range of 0.01-50ngmL(-1) were used for quantification to compensate for matrix effects with the correlation coefficients demonstrating good linearity (0.9940-0.9999). The extraction of the antibiotics from surface waters was performed using solid phase extraction with Oasis HLB cartridges. Accuracy was expressed as recovery with values ranging from 96.52% to 127.30% and from 91.66% to 123.70% for FL and MS detection, respectively.<br /> (Copyright © 2013 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-376X
Volume :
927
Database :
MEDLINE
Journal :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
Publication Type :
Academic Journal
Accession number :
23340484
Full Text :
https://doi.org/10.1016/j.jchromb.2012.12.032