Comparison of TLD calibration methods for 192Ir dosimetry.
- Source :
-
Journal of applied clinical medical physics [J Appl Clin Med Phys] 2013 Jan 07; Vol. 14 (1), pp. 4037. Date of Electronic Publication: 2013 Jan 07. - Publication Year :
- 2013
-
Abstract
- For the purpose of dose measurement using a high-dose rate (192)Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the (192)Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy correction factor was applied to account for the difference in sensitivity of the TLDs in (192)Ir and 6 MV. The results of the four TLD calibration methods are presented in terms of the results of a brachytherapy audit where seven Australian centers irradiated three sets of TLDs in a water phantom. The results were in agreement within estimated uncertainties when the TLDs were calibrated with the (192)Ir source. Calibrating TLDs in a phantom similar to that used for the audit proved to be the most practical method and provided the greatest confidence in measured dose. When calibrated using 6 MV photons, the TLD results were consistently higher than the (192)Ir-calibrated TLDs, suggesting this method does not fully correct for the response of the TLDs when irradiated in the audit phantom.
Details
- Language :
- English
- ISSN :
- 1526-9914
- Volume :
- 14
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of applied clinical medical physics
- Publication Type :
- Academic Journal
- Accession number :
- 23318392
- Full Text :
- https://doi.org/10.1120/jacmp.v14i1.4037