Back to Search Start Over

Comparison of TLD calibration methods for 192Ir dosimetry.

Authors :
Haworth A
Butler DJ
Wilfert L
Ebert MA
Todd SP
Hayton AJ
Kron T
Source :
Journal of applied clinical medical physics [J Appl Clin Med Phys] 2013 Jan 07; Vol. 14 (1), pp. 4037. Date of Electronic Publication: 2013 Jan 07.
Publication Year :
2013

Abstract

For the purpose of dose measurement using a high-dose rate (192)Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the (192)Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy correction factor was applied to account for the difference in sensitivity of the TLDs in (192)Ir and 6 MV. The results of the four TLD calibration methods are presented in terms of the results of a brachytherapy audit where seven Australian centers irradiated three sets of TLDs in a water phantom. The results were in agreement within estimated uncertainties when the TLDs were calibrated with the (192)Ir source. Calibrating TLDs in a phantom similar to that used for the audit proved to be the most practical method and provided the greatest confidence in measured dose. When calibrated using 6 MV photons, the TLD results were consistently higher than the (192)Ir-calibrated TLDs, suggesting this method does not fully correct for the response of the TLDs when irradiated in the audit phantom.

Details

Language :
English
ISSN :
1526-9914
Volume :
14
Issue :
1
Database :
MEDLINE
Journal :
Journal of applied clinical medical physics
Publication Type :
Academic Journal
Accession number :
23318392
Full Text :
https://doi.org/10.1120/jacmp.v14i1.4037