Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents: first in vivo experience.

Authors :
Xia Y
Junjie Y
Ying Z
Bai H
Qi W
Qinhua J
Yundai C
Source :
European journal of radiology [Eur J Radiol] 2013 Apr; Vol. 82 (4), pp. 617-22. Date of Electronic Publication: 2012 Dec 23.
Publication Year :
2013

Abstract

Objective: To investigate the accuracy of 128-slice dual-source CT using high-pitch spiral mode (HPS) for the assessment of coronary stents.<br />Methods: We conducted a prospective study on patients with previous stent implantation due to recurred suspicious symptoms of angina with positive findings at stress testing scheduled for coronary angiography (CA), while dual source computed tomography (DSCT) examinations were randomly done by one of the three different scan modes [HPS, sequential mode (SEQ), low-pitch spiral mode (LPS)] one week before CA examinations. The image quality, radiation dose and stent patency of DSCT were evaluated blinded to the results of CA.<br />Results: 180 patients with total 256 stents were enrolled in this study. There was no significant difference on the image quality of DSCT by HPS (1.4±0.5), SEQ (1.5±0.5) and LPS (1.3±0.6) (P>0.05). The noise of images reconstructed with B26f kernel in HPS is significantly increased than in SEQ/LPS (P<0.05), while no significant difference with images reconstructed with B46f kernel (P>0.05). Heart rate (HR) variability had a slight impact on the image quality for HPS (P<0.05), not for LPS/SEQ (P>0.05). In the assessment of stent restenosis compared with CA on per-stent basis, there was no significant difference on sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of DSCT using HPS (100%, 97.1%, 83.3%, 100%), LPS (92.3%, 95.9%, 80%, 98.6%) and SEQ (93.3%, 97.3%, 87.5%, 98.6%) (P>0.05). The effective dose of DSCT by HPS (1.0±0.5mSv) is significant less than that by SEQ (3.0±1.4mSv) or LPS (13.0±5.4mSv) (P<0.01).<br />Conclusions: DSCT using HPS provides good diagnostic accuracy on coronary stent patency compared with CA, similar to that by SEQ/LPS, whereas with lower effective dose in patients with HR lower than 65bpm.<br /> (Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1872-7727
Volume :
82
Issue :
4
Database :
MEDLINE
Journal :
European journal of radiology
Publication Type :
Academic Journal
Accession number :
23265926
Full Text :
https://doi.org/10.1016/j.ejrad.2012.11.033