Back to Search
Start Over
Intraocular pressure variation during femtosecond laser-assisted cataract surgery using a fluid-filled interface.
- Source :
-
Journal of cataract and refractive surgery [J Cataract Refract Surg] 2013 Jan; Vol. 39 (1), pp. 22-27. - Publication Year :
- 2013
-
Abstract
- Purpose: To evaluate intraocular pressure (IOP) during femtosecond laser-assisted cataract surgery using a fluid-filled interface.<br />Setting: Ruhr University Eye Clinic, Bochum, Germany.<br />Design: Prospective clinical trial.<br />Methods: The absolute IOP was measured with a modified Schiotz tonometer before and after laser-assisted cataract surgery with and without a fluid-filled interface (Liquid Optics interface, Catalys Precision Laser System).<br />Results: The study evaluated 100 eyes. Tonometry and laser cataract surgery were completed successfully in all eyes. The mean preoperative IOP was 15.6 mm Hg ± 2.5 (SD). Upon application of the suction ring and vacuum, the mean IOP rose to 25.9 ± 5.0 mm Hg and remained nearly constant after the laser procedure (27.6 ± 5.5 mm Hg). After removal of the suction ring, the mean IOP was 19.1 ± 4.4 mm Hg. The IOP 1 hour after surgery was not significantly higher than the preoperative values.<br />Conclusions: The results indicate a minor increase in IOP using the fluid-filled interface. This is in strong contrast to substantially higher values reported in the literature with flat and curved applanating contact interfaces.<br /> (Copyright © 2013 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1873-4502
- Volume :
- 39
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of cataract and refractive surgery
- Publication Type :
- Academic Journal
- Accession number :
- 23245360
- Full Text :
- https://doi.org/10.1016/j.jcrs.2012.10.038