Back to Search Start Over

[Positional accuracy and quality assurance of Backup JAWs required for volumetric modulated arc therapy].

Authors :
Tatsumi D
Nakada R
Ienaga A
Yomoda A
Inoue M
Ichida T
Hosono M
Source :
Nihon Hoshasen Gijutsu Gakkai zasshi [Nihon Hoshasen Gijutsu Gakkai Zasshi] 2012; Vol. 68 (11), pp. 1486-91.
Publication Year :
2012

Abstract

The tolerance of the Backup diaphragm (Backup JAW) setting in Elekta linac was specified as 2 mm according to the AAPM TG-142 report. However, the tolerance and the quality assurance procedure for volumetric modulated arc therapy (VMAT) was not provided. This paper describes positional accuracy and quality assurance procedure of the Backup JAWs required for VMAT. It was found that a gap-width error of the Backup JAW by a sliding window test needed to be less than 1.5 mm for prostate VMAT delivery. It was also confirmed that the gap-widths had been maintained with an error of 0.2 mm during the past one year.

Details

Language :
Japanese
ISSN :
1881-4883
Volume :
68
Issue :
11
Database :
MEDLINE
Journal :
Nihon Hoshasen Gijutsu Gakkai zasshi
Publication Type :
Academic Journal
Accession number :
23171770
Full Text :
https://doi.org/10.6009/jjrt.2012_jsrt_68.11.1486