Determination of lead in medicinal plants by high-resolution continuum source graphite furnace atomic absorption spectrometry using direct solid sampling.

Authors :
Figuerêdo Rêgo J
Virgilio A
Nóbrega JA
Gomes Neto JA
Source :
Talanta [Talanta] 2012 Oct 15; Vol. 100, pp. 21-6. Date of Electronic Publication: 2012 Sep 01.
Publication Year :
2012

Abstract

A procedure is proposed for Pb determination in medicinal plants by high-resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GF AAS) using direct solid sampling. Among Pd(NO(3))(2), Pd/Mg(NO(3))(2), NH(4)H(2)PO(4) and the W-coated platform tested as chemical modifiers, Pd(NO(3))(2) presented the best performance. Calibration plots (10-1000 pg Pb) with regression coefficients better than 0.999 were typically obtained. Accuracy was checked for Pb determination in five plant certified reference materials. Results were in agreement with reference values at a 95% confidence level (paired t-test). Medicinal plant samples were analyzed by the proposed procedure and line-source GF AAS using slurry sampling as a comparative technique. The RSD was 10% (n=3) for a sample containing 0.88 μg g(-1) Pb. The limit of quantification (dry mass) was 0.024 μg g(-1). The contents of Pb in medicinal plant samples varied in the 0.30-1.94 μg g(-1) range.<br /> (Copyright © 2012 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-3573
Volume :
100
Database :
MEDLINE
Journal :
Talanta
Publication Type :
Academic Journal
Accession number :
23141306