Back to Search Start Over

Design and fabrication of a COP-based microfluidic chip: chronoamperometric detection of Troponin T.

Authors :
Abad L
Javier del Campo F
Muñoz FX
Fernández LJ
Calavia D
Colom G
Salvador JP
Marco MP
Escamilla-Gómez V
Esteban-Fernández de Ávila B
Campuzano S
Pedrero M
Pingarrón JM
Godino N
Gorkin R 3rd
Ducrée J
Source :
Electrophoresis [Electrophoresis] 2012 Nov; Vol. 33 (21), pp. 3187-94. Date of Electronic Publication: 2012 Oct 12.
Publication Year :
2012

Abstract

This work demonstrates the design and fabrication of an all cyclo-olefin polymer based microfluidic device capable of capturing magnetic beads and performing electrochemical detection in a series of gold electrodes. The size of chip is of a microscope slide and features six independent measuring cells for multianalyte detection purposes. The aim of this work is to show that rapid prototyping techniques can be instrumental in the development of novel bioassays, particularly in clinical diagnosis applications. We show the successful determination of troponin-T, a cardiac disease marker, in the clinically relevant range of 0.05-1.0 ng/mL. This methodology achieves a detection limit of 0.017 ng/mL in PBS solutions, and is capable of detecting less than 1 ng/mL in a 1:50 human serum dilution.<br /> (© 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.)

Details

Language :
English
ISSN :
1522-2683
Volume :
33
Issue :
21
Database :
MEDLINE
Journal :
Electrophoresis
Publication Type :
Academic Journal
Accession number :
23065680
Full Text :
https://doi.org/10.1002/elps.201200225