Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a randomized comparison of different acquisition protocols.
- Source :
-
European radiology [Eur Radiol] 2013 Mar; Vol. 23 (3), pp. 614-22. Date of Electronic Publication: 2012 Oct 07. - Publication Year :
- 2013
-
Abstract
- Objectives: To compare the diagnostic performance and radiation exposure of 128-slice dual-source CT coronary angiography (CTCA) protocols to detect coronary stenosis with more than 50 % lumen obstruction.<br />Methods: We prospectively included 459 symptomatic patients referred for CTCA. Patients were randomized between high-pitch spiral vs. narrow-window sequential CTCA protocols (heart rate below 65 bpm, group A), or between wide-window sequential vs. retrospective spiral protocols (heart rate above 65 bpm, group B). Diagnostic performance of CTCA was compared with quantitative coronary angiography in 267 patients.<br />Results: In group A (231 patients, 146 men, mean heart rate 58 ± 7 bpm), high-pitch spiral CTCA yielded a lower per-segment sensitivity compared to sequential CTCA (89 % vs. 97 %, P = 0.01). Specificity, PPV and NPV were comparable (95 %, 62 %, 99 % vs. 96 %, 73 %, 100 %, P > 0.05) but radiation dose was lower (1.16 ± 0.60 vs. 3.82 ± 1.65 mSv, P < 0.001). In group B (228 patients, 132 men, mean heart rate 75 ± 11 bpm), per-segment sensitivity, specificity, PPV and NPV were comparable (94 %, 95 %, 67 %, 99 % vs. 92 %, 95 %, 66 %, 99 %, P > 0.05). Radiation dose of sequential CTCA was lower compared to retrospective CTCA (6.12 ± 2.58 vs. 8.13 ± 4.52 mSv, P < 0.001). Diagnostic performance was comparable in both groups.<br />Conclusion: Sequential CTCA should be used in patients with regular heart rates using 128-slice dual-source CT, providing optimal diagnostic accuracy with as low as reasonably achievable (ALARA) radiation dose.
- Subjects :
- Coronary Angiography methods
Coronary Artery Disease epidemiology
Female
Humans
Incidence
Male
Middle Aged
Netherlands epidemiology
Prevalence
Reproducibility of Results
Risk Factors
Sensitivity and Specificity
Tomography, X-Ray Computed methods
Coronary Angiography statistics & numerical data
Coronary Artery Disease diagnostic imaging
Radiographic Image Interpretation, Computer-Assisted methods
Radiography, Dual-Energy Scanned Projection methods
Tomography, X-Ray Computed statistics & numerical data
Subjects
Details
- Language :
- English
- ISSN :
- 1432-1084
- Volume :
- 23
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- European radiology
- Publication Type :
- Academic Journal
- Accession number :
- 23052644
- Full Text :
- https://doi.org/10.1007/s00330-012-2663-3