Back to Search
Start Over
Dosimetric verification of radiotherapy treatment planning systems in Serbia: national audit.
- Source :
-
Radiation oncology (London, England) [Radiat Oncol] 2012 Sep 12; Vol. 7, pp. 155. Date of Electronic Publication: 2012 Sep 12. - Publication Year :
- 2012
-
Abstract
- Background: Independent external audits play an important role in quality assurance programme in radiation oncology. The audit supported by the IAEA in Serbia was designed to review the whole chain of activities in 3D conformal radiotherapy (3D-CRT) workflow, from patient data acquisition to treatment planning and dose delivery. The audit was based on the IAEA recommendations and focused on dosimetry part of the treatment planning and delivery processes.<br />Methods: The audit was conducted in three radiotherapy departments of Serbia. An anthropomorphic phantom was scanned with a computed tomography unit (CT) and treatment plans for eight different test cases involving various beam configurations suggested by the IAEA were prepared on local treatment planning systems (TPSs). The phantom was irradiated following the treatment plans for these test cases and doses in specific points were measured with an ionization chamber. The differences between the measured and calculated doses were reported.<br />Results: The measurements were conducted for different photon beam energies and TPS calculation algorithms. The deviation between the measured and calculated values for all test cases made with advanced algorithms were within the agreement criteria, while the larger deviations were observed for simpler algorithms. The number of measurements with results outside the agreement criteria increased with the increase of the beam energy and decreased with TPS calculation algorithm sophistication. Also, a few errors in the basic dosimetry data in TPS were detected and corrected.<br />Conclusions: The audit helped the users to better understand the operational features and limitations of their TPSs and resulted in increased confidence in dose calculation accuracy using TPSs. The audit results indicated the shortcomings of simpler algorithms for the test cases performed and, therefore the transition to more advanced algorithms is highly desirable.
- Subjects :
- Calibration
Carcinoma radiotherapy
Humans
Lung Neoplasms radiotherapy
Multicenter Studies as Topic
Phantoms, Imaging standards
Radiometry methods
Radiometry standards
Radiotherapy Planning, Computer-Assisted standards
Radiotherapy, Conformal methods
Radiotherapy, Conformal standards
Serbia
Algorithms
Clinical Audit
Quality Assurance, Health Care
Radiotherapy Dosage standards
Radiotherapy Planning, Computer-Assisted methods
Subjects
Details
- Language :
- English
- ISSN :
- 1748-717X
- Volume :
- 7
- Database :
- MEDLINE
- Journal :
- Radiation oncology (London, England)
- Publication Type :
- Academic Journal
- Accession number :
- 22971539
- Full Text :
- https://doi.org/10.1186/1748-717X-7-155