Back to Search
Start Over
Validation of the Vivago Wrist-Worn accelerometer in the assessment of physical activity.
- Source :
-
BMC public health [BMC Public Health] 2012 Aug 22; Vol. 12, pp. 690. Date of Electronic Publication: 2012 Aug 22. - Publication Year :
- 2012
-
Abstract
- Background: Most accelerometers are worn around the waist (hip or lower back) to assess habitual physical activity. Wrist-worn accelerometers may be an alternative to the waist-worn monitors and may improve compliance in studies with prolonged wear. The aim of this study was to validate the Vivago® Wrist-Worn Accelerometer at various intensities of physical activity (PA) in adults.<br />Methods: Twenty-one healthy adults aged 20-34 years were recruited for the study. Accelerometer data and oxygen uptake (VO2) were measured at sedentary, light, moderate and vigorous levels of PA.<br />Results: Activity categories and accelerometer counts were: sedentary, 0-15 counts·min-1; light, 16-40 counts·min-1; moderate, 41-85 counts·min-1; and vigorous activity, >; 85 counts·min-1. ANOVA repeated measures was used to determine the relationship between accelerometry data output and oxygen consumption (r = .89; p <; .001). The Bland and Altman method showed good agreement in the assessment of energy expenditure between the indirect calorimetry and the data obtained by the accelerometer.<br />Conclusions: Results of the study suggest that the Vivago® wrist-worn accelerometer is a valid measure of PA at varying levels of intensity. The study has also defined threshold values at 4 intensities and hence te Vivago® accelerometer may be used to quantify PA in free living conditions among adults. This device has possible application in treating a variety of important health concerns.
Details
- Language :
- English
- ISSN :
- 1471-2458
- Volume :
- 12
- Database :
- MEDLINE
- Journal :
- BMC public health
- Publication Type :
- Academic Journal
- Accession number :
- 22913286
- Full Text :
- https://doi.org/10.1186/1471-2458-12-690