A photodiode amplifier system for pulse-by-pulse intensity measurement of an x-ray free electron laser.

Authors :
Kudo T
Tono K
Yabashi M
Togashi T
Sato T
Inubushi Y
Omodani M
Kirihara Y
Matsushita T
Kobayashi K
Yamaga M
Uchiyama S
Hatsui T
Source :
The Review of scientific instruments [Rev Sci Instrum] 2012 Apr; Vol. 83 (4), pp. 043108.
Publication Year :
2012

Abstract

We have developed a single-shot intensity-measurement system using a silicon positive-intrinsic-negative (PIN) photodiode for x-ray pulses from an x-ray free electron laser. A wide dynamic range (10(3)-10(11) photons/pulse) and long distance signal transmission (>100 m) were required for this measurement system. For this purpose, we developed charge-sensitive and shaping amplifiers, which can process charge pulses with a wide dynamic range and variable durations (ns-μs) and charge levels (pC-μC). Output signals from the amplifiers were transmitted to a data acquisition system through a long cable in the form of a differential signal. The x-ray pulse intensities were calculated from the peak values of the signals by a waveform fitting procedure. This system can measure 10(3)-10(9) photons/pulse of ~10 keV x-rays by direct irradiation of a silicon PIN photodiode, and from 10(7)-10(11) photons/pulse by detecting the x-rays scattered by a diamond film using the silicon PIN photodiode. This system gives a relative accuracy of ~10(-3) with a proper gain setting of the amplifiers for each measurement. Using this system, we succeeded in detecting weak light at the developmental phase of the light source, as well as intense light during lasing of the x-ray free electron laser.<br /> (© 2012 American Institute of Physics)

Details

Language :
English
ISSN :
1089-7623
Volume :
83
Issue :
4
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
22559516
Full Text :
https://doi.org/10.1063/1.3701713