Back to Search
Start Over
Performance of a fast digital integrator in on-field magnetic measurements for particle accelerators.
- Source :
-
The Review of scientific instruments [Rev Sci Instrum] 2012 Feb; Vol. 83 (2), pp. 024702. - Publication Year :
- 2012
-
Abstract
- The fast digital integrator has been conceived to face most demanding magnet test requirements with a resolution of 10 ppm, a signal-to-noise ratio of 105 dB at 20 kHz, a time resolution of 50 ns, an offset of 10 ppm, and on-line processing. In this paper, the on-field achievements of the fast digital integrator are assessed by a specific measurement campaign at the European Organization for Nuclear Research (CERN). At first, the architecture and the metrological specifications of the instrument are reported. Then, the recent on-field achievements of (i) ±10 ppm of uncertainty in the measurement of the main field for superconducting magnets characterization, (ii) ±0.02 % of field uncertainty in quality assessment of small-aperture permanent magnets, and (iii) ±0.15 % of drift, in an excitation current measurement of 600 s under cryogenic conditions, are presented and discussed.
Details
- Language :
- English
- ISSN :
- 1089-7623
- Volume :
- 83
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- The Review of scientific instruments
- Publication Type :
- Academic Journal
- Accession number :
- 22380108
- Full Text :
- https://doi.org/10.1063/1.3673000