Back to Search Start Over

A novel respiratory rate estimation method for sound-based wearable monitoring systems.

Authors :
Zhang J
Ser W
Goh DY
Source :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2011; Vol. 2011, pp. 3213-6.
Publication Year :
2011

Abstract

The respiratory rate is a vital sign that can provide important information about the health of a patient, especially that of the respiratory system. The aim of this study is to develop a simple method that can be applied in wearable systems to monitor the respiratory rate automatically and continuously over extended periods of time. In this paper, a novel respiratory rate estimation method is presented to achieve this target. The proposed method has been evaluated in both the open-source data as well as the local-hospital data, and the results are encouraging. The findings of this study revealed strong linear correlation to the reference respiratory rate. The correlation coefficients for the open-source data and the in-hospital data are 0.99 and 0.96 respectively. The standard deviation of the estimation error is less than 7% for both types of data.

Details

Language :
English
ISSN :
2694-0604
Volume :
2011
Database :
MEDLINE
Journal :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Publication Type :
Academic Journal
Accession number :
22255023
Full Text :
https://doi.org/10.1109/IEMBS.2011.6090874