Back to Search
Start Over
A mixed signal ECG processing platform with an adaptive sampling ADC for portable monitoring applications.
- Source :
-
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2011; Vol. 2011, pp. 2196-9. - Publication Year :
- 2011
-
Abstract
- This paper describes a mixed-signal ECG processing platform with an 12-bit ADC architecture that can adapt its sampling rate according to the input signals rate of change. This enables the sampling of ECG signals with significantly reduced data rate without loss of information. The presented adaptive sampling scheme reduces the ADC power consumption, enables the processing of ECG signals with lower power consumption, and reduces the power consumption of the radio while streaming the ECG signals. The test results show that running a CWT-based R peak detection algorithm using the adaptively sampled ECG signals consumes only 45.6 μW and it leads to 36% less overall system power consumption.
- Subjects :
- Equipment Design
Equipment Failure Analysis
Reproducibility of Results
Sensitivity and Specificity
Algorithms
Analog-Digital Conversion
Diagnosis, Computer-Assisted instrumentation
Electroencephalography instrumentation
Signal Processing, Computer-Assisted instrumentation
Telemetry instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 2694-0604
- Volume :
- 2011
- Database :
- MEDLINE
- Journal :
- Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
- Publication Type :
- Academic Journal
- Accession number :
- 22254775
- Full Text :
- https://doi.org/10.1109/IEMBS.2011.6090414