A validated high-performance liquid chromatographic method with diode-array detection for the estimation of xyloketal B in rat plasma.
- Source :
-
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences [J Chromatogr B Analyt Technol Biomed Life Sci] 2012 Feb 15; Vol. 885-886, pp. 24-9. Date of Electronic Publication: 2011 Dec 13. - Publication Year :
- 2012
-
Abstract
- A sensitive and specific HPLC-UV method was developed and validated for the determination of xyloketal B in rat plasma. Following liquid-liquid extraction, the separation was performed using an isocratic mobile phase of methanol-acetonitrile-water (30/30/40, v/v/v) on a Phenomenex C(18) column (4.6mm×250mm, 5μm). The eluent was monitored at 220nm and at a flow rate of 0.8mlmin(-1). A linear curve over the concentration range of 1-128μg/ml (r>0.999) was established. The LLOQ of the method was 1μg/ml. Good precision and accuracy at concentrations of 2.5, 25 and 100μg/ml were obtained. The recovery of xyloketal B in plasma was >87.91%. The validated method was found to be specific, precise and accurate in the study. The analytic method was satisfactorily applied to perform preclinical pharmacokinetic study of xyloketal B in rat plasma.<br /> (Copyright © 2011 Elsevier B.V. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1873-376X
- Volume :
- 885-886
- Database :
- MEDLINE
- Journal :
- Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
- Publication Type :
- Academic Journal
- Accession number :
- 22222627
- Full Text :
- https://doi.org/10.1016/j.jchromb.2011.12.005