Back to Search
Start Over
RF pulse optimization for Bloch-Siegert B ₁⁺ mapping.
- Source :
-
Magnetic resonance in medicine [Magn Reson Med] 2012 Sep; Vol. 68 (3), pp. 857-62. Date of Electronic Publication: 2011 Dec 05. - Publication Year :
- 2012
-
Abstract
- The Bloch-Siegert (B-S) method of B ₁⁺ mapping has been shown to be fast and accurate, yet has high SAR and moderately long TE. These limitations can lengthen scan times and incur signal loss due to B(0) inhomogeneity, particularly at high field. The B-S method relies on applying a band-limited off-resonant B-S radiofrequency pulse to induce a B ₁⁺-dependent frequency-shift for resonant spins. A method for optimizing the B-S radiofrequency pulse is presented here, which maximizes B-S B ₁⁺ measurement sensitivity for a given SAR and T(2) . A 4-ms optimized pulse is shown to have 35% less SAR compared with the conventional 6-ms Fermi pulse while still improving B ₁⁺ map angle-to-noise ratio by 22%. The optimized pulse performance is validated both in phantom and in vivo brain imaging at 7 T.<br /> (Copyright © 2011 Wiley Periodicals, Inc.)
Details
- Language :
- English
- ISSN :
- 1522-2594
- Volume :
- 68
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Magnetic resonance in medicine
- Publication Type :
- Academic Journal
- Accession number :
- 22144397
- Full Text :
- https://doi.org/10.1002/mrm.23271